A Traceable Single-Setup Procedure for Simultaneous Static and Dynamic Humidity Calibration: Demonstration on a Broad Range of 48 Commercial Low-Cost Sensors

Felix Witt and Volker Ebert

Abstract. Low-cost relative humidity sensors (LCrHS) are widely used in science, industry, and consumer applications; however, traceable calibration of individual units and proper dynamic characterization are largely unavailable. This work presents a novel calibration procedure, MetDyHuCal, implemented in our dynH2O test setup. Our approach enables metrologically traceable, static and dynamic humidity calibration within a single setup and during the same measurement campaign. Our setup is based on two absolute optical hygrometers, SEALDH-II and MANTA-LAH as primary, SI-traceable references – both utilizing direct tunable diode laser absorption spectroscopy (dTDLAS). The proposed setup and procedure were applied to 48 LCrHS (9 models from 8 manufacturers). Static calibrations reveal substantial sensor model-level offsets and non-linearities, highlighting the necessity for individual correction of each sensor. Dynamic characterizations yields response times that deviate by −50 % to +81 % from the manufacturer datasheet values. Furthermore, several LCrHS models exhibit a significant, systematic dependence of the measured response time on the absolute humidity level. Together, these results illustrate large performance spreads between nominally identical LCrHS. The reproducibility of the MetDyHuCal procedure was assessed by repeated characterisations of 10 sensors, each measured five times on different days and after being removed and reinstalled in the setup. The reproducibility test for the characteristic time constant τ yielded a median relative standard deviation of 1.0 % (mean 1.3 %). We demonstrate that meaningful, traceable performance figures can be obtained using the proposed standardized procedure. Beyond LCrHS, our approach may also be used to characterize the dynamic behaviour of high-end research instruments, such as HAI, our multi-channel dTDLAS on the research aircraft HALO.

Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors. Views expressed in the text are those of the authors and do not necessarily reflect the views of the publisher.
Share
Felix Witt and Volker Ebert

Status: open (until 26 Dec 2026)

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
Felix Witt and Volker Ebert
Felix Witt and Volker Ebert
Metrics will be available soon.
Latest update: 20 Nov 2026
Download
Short summary
Low-cost humidity sensors are widely deployed, yet individual units lack traceable calibration and dynamic characterization. We present MetDyHuCal, a procedure enabling traceable static and dynamic calibration within a single setup, referenced to SI-traceable laser hygrometers. Applied to 48 sensors, it revealed unit-to-unit variability and response times deviating up to 81 % from datasheet values. High reproducibility (1.0 % median deviation) supports its use as a standardized procedure.