Preprints
https://doi.org/10.5194/egusphere-2026-6171
https://doi.org/10.5194/egusphere-2026-6171
23 Nov 2026
 | 23 Nov 2026
Status: this preprint is open for discussion and under review for Atmospheric Measurement Techniques (AMT).

Spectral responsivity calibration of pyrgeometers at PTB's new IR detector comparator facility

Chantal I. Mueller, Tobias Pohl, Moritz Feierabend, Peter Meindl, and Christian Monte

Abstract. The Physikalisch-Technische Bundesanstalt (PTB) has established a new infrared detector comparator facility enabling SI-traceable spectral responsivity calibrations in the wavelength range of 1.0 μm to 17.7 μm. The facility's potential is demonstrated through the first spectrally resolved calibration of two pyrgeometer types in widespread meteorological use, the CGR4 (Kipp & Zonen) and the IR20 (Hukseflux). Pyrgeometers are the standard instrumentation for ground-based measurements of downward longwave radiation (DLR), a key component of the surface radiation budget. Although spectrally non-uniform responsivity has long been recognized as an inherent property of these instruments, a direct SI-traceable spectrally resolved responsivity calibration of commercial pyrgeometers has, to our knowledge, not been reported. The results reveal that the spectral responsivity of both pyrgeometer types varies substantially across the examined spectral range, underscoring the necessity of spectrally resolved characterization for improving DLR measurement accuracy. Beyond pyrgeometer-based measurements, the facility opens new metrological opportunities across a wide range of infrared applications.

Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors. Views expressed in the text are those of the authors and do not necessarily reflect the views of the publisher.
Share
Chantal I. Mueller, Tobias Pohl, Moritz Feierabend, Peter Meindl, and Christian Monte

Status: open (until 28 Dec 2026)

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
Chantal I. Mueller, Tobias Pohl, Moritz Feierabend, Peter Meindl, and Christian Monte

Data sets

Spectral Responsivity of IR20 and CGR4 Pyrgeometers between 3.9 µm and 17.7 µm Measured at PTB's Infrared Detector Comparator Facility Chantal I. Mueller et al. https://doi.org/10.5281/zenodo.23188199

Chantal I. Mueller, Tobias Pohl, Moritz Feierabend, Peter Meindl, and Christian Monte
Metrics will be available soon.
Latest update: 23 Nov 2026
Download
Short summary
Pyrgeometers measure downward longwave radiation, a key component of Earth's energy budget. We characterized the spectral responsivity of two widely used pyrgeometer types from 3.9 µm to 17.7 µm, traceable to the International System of Units, using a newly developed calibration facility. Both exhibit pronounced wavelength-dependent responsivity, which may contribute to humidity-dependent biases. The results provide a basis for improving the accuracy of longwave radiation measurements.
Share