Identifying and addressing long-term Alphasense OPC-N3 sensor degradation

Tristalee Mangin, Kamaljeet Kaur, Dillon Tang, Dustin Reitstetter, Raven Reitstetter, Aaron Hanson, Yiqing Meng, Alex Mass, Nancy Daher, and Kerry E. Kelly

Abstract. Low-cost optical particle counters are increasingly used for long-term particulate matter (PM) monitoring, but their long-term performance remains poorly understood. This study evaluates the field performance of six Alphasense OPC-N3 sensors colocated with Federal Equivalent Method (FEM) instruments in Salt Lake City, Utah, over deployment periods ranging from 5 to 24 months. It develops seasonal PM10 correction models. However, the performance of the correction models for the three longest-deployed sensors exhibits progressive degradation after approximately 10–12 months, with correction model slopes increasing by up to sevenfold. In contrast, colocated Plantower PMS5003T sensors show no comparable long-term degradation in PM2.5 measurements. Comparison of OPC-N3s deployed for more than 12 months in the field with those for less than 3 months in the field reveals substantial differences in PM10 response, with old-versus-new comparisons yielding an average slope of 0.18, compared with 0.72 for old-versus-old and new-versus-new comparisons. Post-deployment laboratory evaluations identify irreversible corrosion of the OPC-N3’s internal elliptical mirror in the most degraded OPC-N3 sensor, with corresponding laboratory regression slope increases of more than an order of magnitude compared to pre-deployment evaluation. Mirror replacement substantially improves laboratory performance, suggesting that mirror replacement is a potential pathway to recover performance. Additionally, we observe a 56.3% decrease in the OPC-N3 PM2.5-to-PMS5003T PM2.5 ratio within the first year for the three longest-deployed sensors, suggesting that this ratio may provide a practical indicator of OPC-N3 sensor degradation when FEM measurements are unavailable. These results provide the first evidence linking long-term OPC-N3 performance degradation to internal mirror corrosion and demonstrate that sensor performance can decline substantially after approximately one year of deployment.

Competing interests: Kerry E. Kelly has a financial interest in the company Tellus Networked Solutions, Inc., which commercializes solutions for environmental monitoring.

Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors. Views expressed in the text are those of the authors and do not necessarily reflect the views of the publisher.
Share
Tristalee Mangin, Kamaljeet Kaur, Dillon Tang, Dustin Reitstetter, Raven Reitstetter, Aaron Hanson, Yiqing Meng, Alex Mass, Nancy Daher, and Kerry E. Kelly

Status: open (until 02 Dec 2026)

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
Tristalee Mangin, Kamaljeet Kaur, Dillon Tang, Dustin Reitstetter, Raven Reitstetter, Aaron Hanson, Yiqing Meng, Alex Mass, Nancy Daher, and Kerry E. Kelly
Tristalee Mangin, Kamaljeet Kaur, Dillon Tang, Dustin Reitstetter, Raven Reitstetter, Aaron Hanson, Yiqing Meng, Alex Mass, Nancy Daher, and Kerry E. Kelly
Metrics will be available soon.
Latest update: 28 Oct 2026
Download
Short summary
Low-cost air quality sensors can help communities track particle pollution, but it is unclear how well they perform over many years. We find that some OPC-N3 sensors begin to lose accuracy after about one year of use. Corrosion on the mirror inside the sensor likely contributes to this decline, while replacing the mirror improves performance. These findings highlight the need for regular sensor checks and maintenance.